Accuracy
- Uncertainty of the Output Value: 1 % of Full Scale
Ambient Temperature Range
- Operating & Storage: -10° to 60° C (14° to 140° F)
Approvals & Certifications
- ECR 1935/2004
- CE
- CIP Version: FDA
- EU Directives
- PED 2014/68/EU
- UL (C,US): UL 61010-1 + CAN/CSA C22.2 No.61010-1
Consumption
- Current: ≤ 25 mA (At 12 Vdc & without the Consumption of the 4-20 mA Output)
Diameter
- Connection Cable: 0.12” to 0.25” (3 to 6.5 mm)
Electrical Connectors
- 2 Outputs Meter (3-Wire): 1 x 5-Pin M12 Male Fixed Connector
- 4 Outputs Meter (3-Wire): 1 x 5-Pin M12 Male & 1 x 5-Pin M12 Female Fixed Connectors
- Maximum
- Shielded Cable
Linearity
- Conductivity Measurement: ± 2 %
Loop Impedance
- 1,100 ohms at 36 Vdc, Max.
- 100 ohms at 12 Vdc, Max.
- 610 ohms at 24 Vdc, Max.
Materials of Construction
- Display: Polycarbonate (PC)
- Navigation Key: PBT
- With CIP Version (M12 Fixed Connectors), Fixed Connector Holder: 316L Stainless Steel
- With Standard Version (M12 Fixed Connectors): Brass Nickel-Plated
Measuring Error
- Conductivity Measurement Deviation: ± (2 % of the Measured Value + 5 µS/cm)
- Temperature Measurement Deviation: ± 1° C (1.8° F)
Measuring Range
- Conductivity Measurement: 10 to 2,000,000 µS/cm
Nut & Bolt Material
- With Standard Version (Nut): Polycarbonate (PC)
Output
- 4-20 mA
- Transistor (NPN): 1-36 Vdc
- Transistor (PNP & NPN): 700 mA - Max.
Pipe Connection
- DN 100 Pipe
- DN 125 Pipe
- DN 15 Pipe
- DN 150 Pipe
- DN 20 Pipe
- DN 200 Pipe
- DN 25 Pipe
- DN 32 Pipe
- DN 40 Pipe
- DN 50 Pipe
- DN 65 Pipe
- DN 80 Pipe
Pressure Range
- PEEK - psi: 0 to 145 (kPa: 0 to 1,000, bar: 0 to 10)
- PP, PVDF - psi: 0 to 87 (kPa: 0 to 600, bar: 0 to 6)
Process Connection Material
- With CIP Version: 316L Stainless Steel
Process Temperature
- PEEK: -15° to 130° C (5° to 266° F)
- PP: 0° to 80° C (32° to 176° F)
- PVDF: -15° to 100° C (5° to 212° F)
Repeatability
- Conductivity Measurement: ± (0.2 % of the Measured Value + 2 µS/cm)
Resolution
- Conductivity Measurement: 0.1 µS/cm
- Temperature Measurement: 0.1° C (0.18° F)
Response Time
- 10 to 90 %: 150 milliseconds
- Conductivity Measurement (t90): 3 to 40 seconds
Reverse Polarity Protection
- Protected Against Reverse Polarity
Seat / Seal Material
- Seal: Silicone
- With Standard Version & with CIP Version (Seal): EPDM
- With Standard Version & with CIP Version (Seal): FKM
Supply Voltage
- 12-36 Vdc ± 10 % Oscillation Rate, Filtered & Regulated
Temperature Range
- Measuring Range: -15° to 130° C (5° to 266° F)
Tri-Clamp Connection
- 1-1/2” Tri-Clamp
- CIP Version: 2” Tri-Clamp
Wetted Materials
- With CIP Version (Sensor Holder): 316L Stainless Steel
- With Standard Version & with CIP Version (Sensor Holder): PEEK
- With Standard Version & with CIP Version (Sensor Holder): PVDF
- With Standard Version (Sensor Holder): Polypropylene (PP)