8222 Conductivity Meter

Brand: Burkert Fluid Control Systems
Does not ship to VA, United States

The Burkert compact meter Type 8222 is designed for measuring the conductivity of fluids. The conductivity meter consists of a sensor, plugged-in and pined to the transmitter (with removable display). 

The sensor comprises a cell with two electrodes and a Pt1000 temperature probe. The sensor itself is available with three different cell constants C, these with C = 0.01 or 0.1 are fitted with stainless steel electrodes and those with C = 1.0 are fitted with graphite electrodes. 

The conductivity meter can operate independent of the display but it will be required for programming the device and also for visualizing continuously the measured and processed data.
Features
  • Perfect for clean water and slightly concentrated liquids
  • Integral device for direct connection to PLC
  • Simulation of process values for diagnostics
  • Three cell constants to cover a wide application range e.g. reverse osmosis

Specifications

Ambient Temperature Range

  • Operating & Storage: -10° to 60° C (14° to 140° F)

Approvals & Certifications

  • Calibration Certificates
  • CE
  • EU Directives
  • FDA
  • PED 2014/68/EU
  • UL (C,US): UL 61010-1 + CAN/CSA C22.2 No.61010-1

BSP Connection

  • G 1-1/2” Male

Circuit Protection

  • Short Circuit: Protected for Transistor Outputs

Consumption

  • Current [3 Outputs Meter (2-Wire)]: ≤ 25 mA at 14 Vdc without Transistors Load, with Current Loop
  • Current [4 Outputs Meter (3-Wire)]: ≤ 5 mA at 12 Vdc without Transistors Load, without Current Loop
  • Current: ≤ 1 A with the 2 Transistors Loads

Cover Material

  • Polycarbonate (PC)

Current

  • 4-20 mA

Diameter

  • Pipe: 0.9 to 4.3” (25 to 110 mm)

Electrical Connectors

  • 3 Outputs Meter (2-Wire): 1 x 5-Pin M12 Male Fixed Connector
  • 4 Outputs Meter (3-Wire): 1 x 5-Pin M12 Male + 1 x 5-Pin M12 Female Fixed Connectors
  • Shielded Cable

Electrical Protection

  • Voltage Peak: Protected

Enclosure, Body Material

  • 316L Stainless Steel, 1.4404
  • PPS

Environmental Protection

  • IP65
  • IP67

Humidity

  • ≤ 85 % Relative Humidity, Non-Condensing

Loop Impedance

  • 1,100 ohms at 36 Vdc, Max.
  • 3 Outputs Meter (2-Wire): 180 ohms at 14 Vdc, Max.
  • 4 Outputs Meter (3-Wire): 100 ohms at 12 Vdc, Max.
  • 610 ohms at 24 Vdc, Max.

Materials of Construction

  • Display: Polycarbonate (PC)
  • Fixed Connector Mounting Plate: 316L Stainless Steel, 1.4404
  • Fixed Connector: Brass Nickel-Plated
  • Navigation Key: PBT

Measuring Error

  • Conductivity Measurement Deviation: ± 3 % of Measured Value
  • Temperature Measurement Deviation: ± 1° C (1.8° F)

Measuring Range

  • Conductivity Measurement: 0.05 to 10,000 µS/cm

Media

  • Water

Nut & Bolt Material

  • Nut: PVC
  • Nut: PVDF

Output

  • 4-20 mA
  • Transistor (NPN): 0.2-36 Vdc

Output Format

  • Open Collector (700 mA, 0.5 A) - Max.

Power Supply

  • 3 Outputs Meter (2-Wire): 14-36 Vdc, Filtered & Regulated
  • 4 Outputs Meter (3-Wire): 12-36, Filtered & Regulated

Pressure Range

  • psi: 0 to 232 (kPa: 0 to 1,600, bar: 0 to 16) - Max.

Process Temperature

  • With G 1-1/2” PVC Nut Connection: 0° to 50° C (32° to 122° F)
  • With G 1-1/2” PVDF Nut Connection: -20° to 100° C (-4° to 212° F) Restricted by the used Adaptor

Resolution

  • Conductivity Measurement: 1 NS/cm
  • Temperature Measurement (Internal Resolution): 0.1° C (0.18° F)

Response Time

  • 10 to 90 %: 150 milliseconds

Reverse Polarity Protection

  • Protected Against Reverse Polarity

RTD Type

  • Pt1000

Screw Material

  • Stainless Steel

Seat / Seal Material

  • Seal: EPDM
  • Seal: Silicone

Sensor, Probe Material

  • Conductivity Sensor: 316Ti Stainless Steel 1.4571
  • Conductivity Sensor: PVDF

Switch Type

  • Transistor (PNP or NPN)

Temperature Range

  • Minimal Temperature Range: 10° to 20° C (50° to 68° F) Corresponding to 4-20 mA
  • Temperature Measurement: -20° to 100° C (-4° to 212° F)

Wetted Materials

  • Electrode: 316Ti Stainless Steel 1.4571 for Cell Constant C = 0.01 or C = 0.1
  • Electrode: Graphite for Cell Constant C = 1.0

show more