Model AFH2 Air Flow Hood

Brand: Dwyer Instruments, LLC

The MODEL AFH2 Air Flow Hood maintains a running average of volumetric air flow measurements and provides the ability to manually record the measurements with the manometer over a given period of time. 

The highly accurate digital manometer has an auto-zero function and will auto-zero at user-defined intervals.

In addition to these features, the unit will alert you when its battery is running low in order to avoid recording inaccurate measurements.
Features
  • Highly Durable and Easy Storage in Included Travel Case
  • No Tools Required to Set up
  • Digital manometer easily mounts to the flow hood or can be used as stand alone instrument

Specifications

Accuracy

  • ± 1 Count (± 0.05 Pascal Typical, ± 0.0002” WC) - Zero System Accuracy
  • ± 3 % of Reading ± 9 cfm (± 4 L/sec, 14 m³/hour) - Supply & Exhaust

Approvals & Certifications

  • CE

Communication

  • Baud Rate: 9600 bit/sec

Dimensions

  • 2” x 2” (600 mm x 600 mm) - Hood Only
  • 30” x 24” x 24” (965 x 610 x 610 mm)

Display

  • 0.375” High Digits

Flow Capacity

  • 25 to 1,176 scfm (43 to 2,000 m³/hour) - Supply
  • 29 to 1,176 scfm (49 to 2,000 m³/hour) - Exhaust

Media

  • Air

Operating Temperature

  • 0° to 50° C (32° to 122° F)

Output

  • RS-232

Power Requirement

  • 8.4 V NiMH Battery
  • 9 V Alkaline Battery

Pressure Range

  • Inches WC: 60 - Max.

Response Time

  • 2 Minute - Warm-Up

Rotation

  • Any 45° Change 0.0004” W.c. (0.1 Pascal) Typical

Span & Zero Adjustments

  • Better than 0.1 % of Range in Use Per 2° F (1° C)

Stability

  • 0.1 % of Range Per 1° C (2° F)

Storage Temperature

  • -5° to 50° C (23° to 122° F)

Temperature Effect

  • ± 0.1 % of Range in Use Per 2° F (1° C)

  • Measure Volumetric Air Flow from Diffusers, Grilles & Registers

show more